Measurement Method | Electromagnetic induction or eddy-current *Depends on probe |
---|
Applications | Non-magnetic coating on ferrous substrate or insulating coating on non-ferrous substrate *Depends on probe |
---|
Measurement Range and Precision | Depends on probe. See "Applicable probe" column. |
---|
Resolution | 0.1µm for less than 100μm, 1μm for 100μm or more (when displayed in μm)0.01mm for less than 10mm, 0.1mm for 10mm or more (when displayed in mm) |
---|
Conformity Standard | Electromagnetic induction: JIS K5600-1-7, JIS H0401, JIS H8401, JIS H8501, / ISO 2808, ISO 2064, ISO 1460, ISO 2178, ISO 19840 / ASTM B 499, ASTM D 7091, ASTM E 376 Eddy-current: JIS K5600-1-7, JIS H8680-2, JIS H8501 / ISO 2808, ISO 2360, ISO 2064, ISO 19840 /ASTM B 244, ASTM D 7091, ASTM E 376 |
---|
Applicable Probe | EP-100, HP-100, EP-110, HP-110 |
---|
Statistical Functions | Number of measurements, Average value, Standard deviation, Maximum value, Minimum value, Trend graph |
---|
Display | Digital (Blanview🄬 2.7 inch full color LCD) |
---|
Data memory | L-600 Standard model: 50,000 items L-600 Expert model: 100,000 items |
---|
Data Output | L-600 Standard model: USB-C L-600 Expert model: USB-C, Bluetooth |
---|
Power Source | 1.5V (AA alkaline battery) ×4 |
---|
Dimensions & Weight | 70 (W)×155 (D)×35 (H) mm, 0.2 kg |
---|
Accessories | Iron substrate (for Fe probe) Aluminium substrate (for NFe probe) Calibration foils x 4 (10, 100, 500, 1000μm) Battery 1.5V (AA alkaline) x 4 Probe adapter Carrying pouch Strap USB cable (Type-C) Connector cover List of compatible probes & optional calibration foils Operating Manual |
---|
Options | Various Fe/NFe probes (required option) Calibration foils (thickness other than accessories) Measuring Stand LW-990 Soft carrying case Bluetooth communication dongle VZC76 (Only for expert model) Calibration certificate (calibration certificate, calibration record, traceability flow chart) |
---|
PC Software | LDL Export |